Start at Step 1 if the delivered system card is already in the Eagle R508. Use this guide to make one short IQ recording from the browser and confirm that the device, receiving server and capture volume are working.
| What is ready? | Where to start |
|---|---|
| R508 and receiving server are already set up | Step 1 — connect and sign in |
| R508 has its system card, and the receiving server is new | Configure the receiving server |
| R508 needs a new system card or recovery | Prepare a system card |
| R508 boots and needs a software update | Verify and install a signed update |
Have the R508, the power supply and connection drawing for your hardware revision, a prepared system card, a management Ethernet cable and the matching 100G cable. Prepare a management computer and a receiving server. The delivery record gives the device identity, how to find its management address, the intended application, host packages and data network addresses.
For the first link and file check, follow the delivery instructions for the RF connectors and leave DAC output disabled. Before connecting an external source, check the input limits for your own hardware. Connectors and power
http://DEVICE_MANAGEMENT_IP:8080/, replacing the placeholder with the assigned address. For a delivery image with fallback support and no DHCP, follow the direct connection steps before trying http://10.168.1.99:8080/. The fallback requires a successful address conflict check. Allow time for first-boot initialization.Ready when ULMUS Device Overview shows your device identity, STREAM as the active application and a ready state. Then prepare the data path and receiving server.
Check the dedicated 100G port, the mounted capture volume, and the Device SDK and DAQ Server specified in your delivery. For a new server, complete the Ubuntu server setup walkthrough and return here. Install the software on the system disk and save IQ on the capture volume.
| Address to open | What you do there |
|---|---|
| ULMUS Web on the R508 | Verify the device, active app and ready status, and create device API credentials as instructed |
| DAQ Web on the receiving server | Start, monitor and stop your own recording, and find saved tasks |
| Desktop application | Use the package in your delivery to connect to the same DAQ service, browse files and analyze them |
Use DAQ Web to start and stop this recording. When other clients are connected, the client that owns the task should control it.
server.status and device.status in the status or method view. Confirm the service is idle, the device identity is correct, STREAM is ready, the 100G link is up and the capture directory is writable. DAQ Web interfaceacquisition.start and paste the JSON into its parameter field. Choose a new task name. This example caps the task at one 128 MiB segment. Check that the active DAQ manifest accepts the fields.acquisition.stop with {}.| Field | First recording example |
|---|---|
| Bandwidth / IQ sample rate per channel | 8 MHz / 10 MSPS, BW8M_10MSPS |
| Channels | ADC1–8. Each adc_channels entry uses zero-based numbering |
| NCO | 500 MHz, entered as 500000000 Hz. Use the signal's setting when using an external source |
| Packet header and receive streams | V1 header, one UDP stream |
| Recording limit | One 128 MiB segment |
{
"task_id": "r508_first_01",
"profile": "BW8M_10MSPS",
"adc_channels": [0, 1, 2, 3, 4, 5, 6, 7],
"center_frequency_hz": 500000000,
"packet_header_enabled": true,
"flow_count": 1,
"segment_size_mib": 128,
"segment_count": 1
}
Open File Management in DAQ Web and find your task. You can also open the capture directory named in the deployment record on the receiving server and find the subdirectory named for your task.
| Check | Expected result |
|---|---|
| Task | This task has finished and DAQ is idle. The name matches the one you submitted |
| Files | There is a task.json file and at least one .bin segment with a positive valid byte count |
| Content | The metadata agrees with the selected sample rate, channels and V1 packet mode. valid_bytes fits within the actual length of its segment |
| Continuity | The task adds no NIC drops, receive errors or sequence gaps. Check sequence numbers using the V1 header and skip any preallocated tail beyond the valid length |
The first recording is complete. For waveform or phase analysis, open this committed file and use its metadata to select the sample rate, channels and interval. File and IQ calibration examples
| What you see | Next check |
|---|---|
| ULMUS will not open | Check the management cable, DHCP lease or fallback address, and port :8080. Management network |
| ULMUS works and DAQ reports the device offline | Check whether the receiver can reach the device management address, and verify the device address, API credential and application match in DAQ |
| Task starts with zero received packets | Check both QSFP28 link states, the data subnet, MTU 9000, the configured destination and the receive port |
| Packets arrive and no file is saved | Check the mount, service-account write permission, free space and whether commit completed after stopping |
| The request is rejected | Use stream.profile.list and the active DAQ manifest to check the channel group, header, stream count and segment size |
For support, note the task ID, exact error, device and host versions, link status and file summary. Troubleshooting
| What to do | Guide |
|---|---|
| Increase bandwidth, channels or stream count | STREAM profiles, channels and packet format |
| Use the desktop application or build your own | Use the desktop package in your delivery, or the C++ / Python and direct RPC examples |
| Save RAW payload without an application header | RAW |
| Take spectrum snapshots or operate DAC | SAM |
Confirm that your own capture has ended, the files are committed and RF output has been shut down according to your delivery procedure. Follow the normal shutdown steps for the device and server if they need to be powered down. Keep the recording directory and its task metadata.