Stop the current task and note the application's error before troubleshooting. The checks below do not require ordinary users to change registers, USB firmware settings or low-level services. Administrator steps are identified separately.
| Symptom |
Expected state |
Check |
| Device IP unknown or application cannot connect |
Actual H102 IP obtained; application connected with configuration readback |
Network and connection |
| Profile unchanged or preparation takes time |
Preparation finished; actual settings match the selection |
Profiles and preparation |
| Reception does not start, no IQ, or USB unavailable |
Received sample counts increase without task errors |
Reception and USB |
| Data arrives but the expected signal is missing |
Port, channel, frequency and plot range match the signal |
Spectrum and RF signals |
| Recording fails or data is interrupted |
Normal completion with file and continuity checks passed |
Recording and integrity |
| Stop remains pending or recovery fails |
Stop/recovery completed; device can be released |
Stop and retry |
Use matching IIO GUI or gr-logicrf software for ordinary operation. A Linux password is not an application-connection workaround. See matching versions.
¶ Network and connection
- Confirm that H102 has booted and is connected to the intended Ethernet network. Ethernet lights indicate a physical link, not completed system startup or working application services.
- Find this device's current IP in the router's DHCP list, or ask the administrator to query it through the console. A new card or board can receive a different address; do not reuse another board's address.
- In the GUI, expand Connection, enter the address in IP / host, then select Connect / refresh. The computer must be able to reach H102. Ask the network administrator to check routing and access rules across subnets.
- If connection still fails, save the error and IP used. Have the administrator check device control on TCP 30432, IIO on TCP 30431, and matching device services. Do not disable authentication checks, broadly change firewall rules or bypass version checks with a default password.
Expected result: the application is connected and reads the active profile, sample rate, RF bandwidth and center frequency. A successful ping alone does not establish application readiness; on networks that block ping, a failed ping alone does not establish that the device is offline.
¶ Profiles and preparation
- A profile dropdown shows a pending selection. Stop reception, recording and transmission, select a built-in profile and choose Apply profile. Wait for actual settings to update.
- Changing sample rate/RF bandwidth reinitializes the radio and can take more than 20 seconds. Wait for completion or an explicit error rather than repeatedly pressing Start or Apply. Same-profile tuning time is not a prediction of profile-change time.
- If the device is busy, stop and disconnect the application using it. A paused GNU Radio panel may still hold the device; release it or close that flowgraph.
- Shared LO mode requires matching RX/TX carriers. Do not request conflicting frequencies on channels sharing an LO.
- After an apply error, read actual settings again rather than assuming the dropdown is active. Save the error and follow Profiles. Have an administrator resolve version mismatches; do not edit compatibility checks.
Expected result: preparation/apply is no longer pending, actual profile/rate/bandwidth match the request, and acquisition is available.
¶ Reception and USB
- Confirm that the device is connected, the profile has been applied and the intended RX channel is selected.
- Start with 4 MSPS / 2 MHz, RX1, Normal IQ. Reduce rate or channel count if demand exceeds the budget; changing only the plot rate does not reduce hardware traffic.
- In the IIO GUI, select live reception and the RX play icon. In GNU Radio, generate/run the flowgraph, then start acquisition in its panel. Opening or generating a flowgraph alone does not acquire data.
- Check whether sample counts increase and whether the application reports a connection, timeout or busy error. An unconnected RF input can show noise, but it should not make received byte counts remain zero.
- If there is still no data, follow Stop and retry, preserving the error. Do not open a second client to take over an active device.
- The IIO GUI receives IQ over Ethernet only. For FT601 USB, use a GNU Radio flowgraph with gr-logicrf and a prepared USB backend.
- Connect H102's USB3.0 / FT601 data port to the computer running the flowgraph. The UART/USB2 console port is not an IQ port. Keep Ethernet connected for device control.
- Use a SuperSpeed data cable and USB3 host port, preferably a direct connection. Stop transfers and release the device before changing cables or ports.
- An administrator must install the local backend, FTDI library and user permissions as described in USB reception. A cable alone does not prepare a missing backend or local service.
- On Linux, 5000M in
lsusb -t is the negotiated USB link speed, not proof of received IQ or lossless operation. Acquisition still needs the matching backend, selected device and valid settings. If it shows 480M, check the data port, cable and host connection while stopped.
Expected result: USB acquisition starts, actual IQ counts increase, no transfer error is reported, and the task stops normally. A USB device name or speed alone does not meet this check.
¶ Spectrum and RF signals
- Select Normal IQ. Digital RAMP16 is a path-test source, not received RF.
- Check channel and antenna connections. Built-in profiles use RX1A/RX2A; choosing RX1/RX2 does not switch to RX1B/RX2B. See Hardware for locations.
- Check actual device frequency. The signal must fall within the current RF bandwidth; changing an axis label alone does not tune hardware.
- Reset spectrum zoom and waterfall limits. Noise without a signal source is normal; valid IQ does not guarantee a signal exists in the selected band.
- If the input is saturated, levels are unexpected or loopback wiring is unclear, stop TX first and check the approved signal-level and attenuation plan. Never connect TX directly to RX or blindly increase TX power to find a signal.
Spectrum dBFS is a digital relative level, not directly dBm. Dual reception does not automatically provide equal phase or a synchronized starting sample.
¶ Recording and integrity
- A snapshot is not a whole-session recording. The GUI save icon exports the latest block. The round record icon starts a new complete recording session. See Recording.
- Budget both transport and storage: IQ16 uses sample rate × channels × 4 bytes; GNU Radio cf32 files use sample rate × channels × 8 bytes. Dual RX cf32 at 8 MSPS needs about 128 MB/s of disk writes. See Data operation.
- Use host storage with enough free space and sustained speed. A short write precheck does not replace checks over the recording window; competing disk work, full storage or transport errors can still fail a task.
- At completion, check the application result and
recording.json. GUI PASS_ADC_DMA_TO_FILE_NO_DROP with absolute_continuity_proven: true applies to that recording window. GNU Radio uses its corresponding completed manifest and integrity result. A file's existence or a moving plot is insufficient.
- Failed recordings retain
.partial files. Do not rename them as complete, pad them with zeros or treat a later successful retry as proof of the earlier interval. Preserve the error, reduce load and start a new recording.
GUI .iq/.bin uses IQ16; GNU Radio example .cf32 uses complex float32. Reading the wrong format causes misleading plots. Network IIO does not include the native USB GNSS header and must not be decoded using the same time format. See File formats and time.
- Select Stop in the active application. Use the appropriate RX or TX control in the GUI. Allow recordings to finish writing and checking; wait for completion in GNU Radio panels.
- After normal task completion, disconnect the GUI or release the device/close the GNU Radio flowgraph. Check that another application is not still using it.
- Save the error and task directory, reconnect and read back actual settings. Try lower-rate single-channel reception before restoring the original task.
- If recovery is unconfirmed, stopping remains pending or the device cannot be released, do not unplug USB, delete lock files, force controller resets or repeatedly restart services. Leave the device out of use and contact the administrator or Support. An administrator should check its state before arranging a normal system shutdown/restart.
Disconnecting a host application is not the same as shutting down H102. Before removing power, taking out the card or changing wiring, stop tasks and have the administrator shut H102 down normally. Wait for shutdown before removing power.
Run these commands on H102 through its console or SSH. Ordinary GUI reception does not require running them each time:
H102 shell · Run after SSH or serial login. Use the interface and configuration names on this unit.
ip -4 address show dev eth0
uname -r
sudo h102-config-sdk1 status
systemctl --failed --no-pager
- The IP output should show the actual device IPv4 address.
- Kernel and profile information should match the software versions, and actual settings should agree with the application.
- If
systemctl --failed lists a failed service, keep its name and error. Check current system services rather than disabling services.
For startup that never completes, check the image source, full checksum and write verification, then follow Quick start for console checks. Rewriting the card overwrites user data: back up first and do not use reimaging as the first response to an ordinary connection error. If an SSH host identity changes, have an administrator verify it through a trusted console; retain host-key checking.
| Item |
What to provide |
| Device |
Hawk H102 model, hardware revision and system image version |
| Application |
IIO GUI or gr-logicrf version, computer OS, Ethernet or USB path |
| Task |
Profile ID, RX/TX channels, center frequency, connections and ordered reproduction steps |
| Symptom |
Expected/actual result, error text or screenshot, and approximate time |
| Task records |
Keep any directory shown by the application; initially provide the result file and location, with further logs requested as needed |
The GUI defaults to ~/h102-iio-captures; use the exact task location shown in the application. Contact LogicRF through Support, with the product and short issue summary in the subject. Do not send passwords, tokens, SSH private keys or IQ content without authorization. You do not need to collect every low-level log before asking for help.
Use system checks and the new-device checker to inspect the services included in Kuiper 1.2.1. Record failed units and the actual version before maintenance.
For r3-written cards without a DHCP lease, use the configured fallback (default 10.168.1.199/24) and a free PC address in the same subnet. See automatic networking.